A four-gas monitor that draws its sample down a hose. Everything that makes this instrument different from a clip-on diffusion detector comes from that one fact.
Minimum 980 × 600 px. Recommended 1920 × 1080 (16:9) or larger. Below the minimum the simulator shows a "Screen Too Small" notice instead of running.
A diffusion detector sits in the atmosphere it measures. An aspirated instrument does not — it stands at the manhole while a pump pulls the atmosphere up a hose to it. Three consequences, and they are the whole syllabus:
| Channel | Range | Resolution | T90 | 1st alarm | 2nd alarm | TWA | STEL | Over |
|---|---|---|---|---|---|---|---|---|
| Combustible (CH4) | 0–100 %LEL | 1 %LEL | <30 s | 10 %LEL | 50 %LEL | — | — | 100 %LEL |
| Oxygen (O2) | 0–25 vol% | 0.1 vol% | <20 s | 19.5 vol% low | 23.5 vol% high | — | — | 40 vol% |
| Carbon monoxide | 0–150 ppm | 1 ppm | <30 s | 25 ppm | 50 ppm | 25 ppm | 200 ppm | 500 ppm |
| Hydrogen sulfide | 0–30 ppm | 0.5 ppm | <30 s | 5.0 ppm | 30.0 ppm | 10.0 ppm | 15.0 ppm | 100 ppm |
Oxygen is the exception to every rule here: its first alarm is a LOW alarm and its second is a HIGH alarm, because both depletion and enrichment are hazards.
| Switch | Short press | Hold |
|---|---|---|
| ▲ / AIR | Step up through a menu | Run the fresh air adjustment |
| ▼ / RESET | Reset alarms, or step down in a menu | — |
| DISPLAY | Move between detection mode and the display/setting menu | — |
| POWER / ENTER | Confirm a menu selection | 3 s — switch the instrument on or off |
Four quadrants — combustible and oxygen on the top row, carbon monoxide and hydrogen sulfide beneath —
each with a value and a bar graph. The status strip carries a pilot indicator that blinks while detecting,
a pump indicator that turns only while gas is genuinely moving, the battery icon and the clock. In an
alarm the affected value blinks; over-range shows ∩∩∩ and a failed sensor shows
---. Neither is ever a number you can act on.
DISPLAY opens it; ▲/▼ step through PEAK, STEL, TWA, ALARM-P, PUMP and ID. PEAK holds the highest reading seen since power-on — except oxygen, where it holds the lowest. PUMP with ENTER stops and starts the suction.
The left panel computes it live:
transit (s) = hose length (m) × hose volume per metre (mL/m) ÷ pump flow (mL/s)
With a 0.75 L/min pump and a typical sampling hose that is roughly 1.4 seconds for every metre. A 20 m hose is close to 30 seconds before the atmosphere at the probe even reaches the sensor — and then the sensor still needs its own T90 on top.
The SAMPLE LINE bar above the instrument fills as the column is exchanged. While it is amber you are looking at hose air. When it turns green the reading represents the probe.
Hold POWER for three seconds. The instrument lights every segment, then walks its date/time, battery, gas names, full scales, first and second alarm points, STEL, TWA and ID before entering detection mode. Watch that scroll — it is how you confirm the setpoints are what you expect.
In genuinely clean air, hold AIR. The instrument purges, then zeroes the combustible and toxic channels and sets oxygen to 20.9 vol%.
FAIL CAL — try it deliberately in the deckhead
scenario and watch it refuse.Gas alarms latch. Once a setpoint is reached the alarm stays up even after the gas clears. RESET clears it only when the reading has genuinely returned below the setpoint — pressing it while still in the gas is refused, and the simulator tells you so. TWA and STEL alarms never reset: the exposure has already happened.
Fault alarms auto-recover once the cause is gone, except the flow fault, which needs the restriction removed and then a deliberate RESET.
Objective: complete the pre-start checks, adjust in fresh air, survey all three levels waiting the full transit each time. Expected result: all four channels within limits at every level; entry may be declared safe. Pass: 70. Par: 12 min.
Objective: recognise that the manhole reading does not describe the space. Oxygen falls from 20.9 vol% at the top to 13.4 vol% at the bottom and H2S rises to 38 ppm. Expected result: the top reads clean; only a bottom sample finds the hazard. The correct answer is to refuse entry. Pass: 80. Par: 15 min.
Objective: the mirror image — 62 %LEL at the top falling to zero at the bottom. A trainee who tests only the bilge declares a flammable space safe. Observe both alarm steps and practise resetting the latch only after withdrawing. Pass: 80. Par: 12 min.
Objective: the probe filter is clogged from the start. Suction falls below threshold,
FAIL LOW FLOW raises and the reading stops tracking the probe.
Expected result: notice the fault, treat the reading as void, clear the filter from the
pre-start checks panel, press RESET, and re-sample.
Pass: 80. Par: 10 min.
Ctrl+Shift+F opens the fault panel. Available faults: clogged probe filter, split hose, degraded pump, H2S sensor failure, CO zero drift (+18 ppm) and low battery. Each produces a realistic consequence rather than a binary break — the split hose, for example, dilutes the sample toward clean air rather than simply stopping it.
| Key | Action |
|---|---|
| F | Fullscreen |
| Space | Pause / resume |
| Esc | Close a modal |
| ? | Shortcut help |
| Ctrl+Shift+F | Fault injection panel |
| Ctrl+Shift+T | Trend plots |
| Ctrl+Shift+S | Score display |
Start at 100.
| Event | Change |
|---|---|
| Level logged before the hose transit completed | −5 |
| Alarm left unacknowledged over 30 s | −2 |
| Incorrect entry decision | −25 |
| Phase checklist completed | +5 |
| Correct entry decision | +15 |
Grades: A 90+, B 75–89, C 60–74, D below 60.
| Symptom | Cause | Action |
|---|---|---|
| Reading will not change when the probe moves | No suction, or a blocked line | Check the pump indicator and the suction flow gauge; inspect the probe filter |
FAIL LOW FLOW | Restricted probe, hose or filter | Remove the cause, then press RESET |
FAIL CAL | Fresh air adjustment attempted in contaminated air | Withdraw to genuinely clean air and repeat |
| RESET does nothing | The gas is still above the setpoint | Withdraw and let the reading recover first — this is correct behaviour |
Channel shows --- | Sensor failure | That gas cannot be detected. The instrument is not fit for entry testing |
Reading shows ∩∩∩ | Above the service range | Withdraw immediately; the true value is unknown and higher than displayed |