No pump, no hose. This instrument measures the air touching its own sensor face, which is the air you are standing in. Everything below follows from that.
Minimum 980 × 600 px. Recommended 1920 × 1080 (16:9) or larger.
This simulator has a sibling, the Pump-Aspirated Confined Space Gas Monitor. They are a matched pair and are best taught together, because the difference between them is a safety decision, not a product preference.
| Aspirated | Diffusion (this one) | |
|---|---|---|
| How gas reaches the sensor | A pump pulls it down a hose | It arrives on its own, by diffusion |
| Can it test a space you are not in? | Yes — lower the probe from outside | No — it only knows where it is |
| The delay you must respect | Hose transit time, then sensor T90 | Sensor T90, plus whatever your clothing adds |
| Primary role | Pre-entry survey of a space | Personal exposure monitoring and evacuation alarm |
| Channel | Range | Resolution | 1st alarm | 2nd alarm | TWA | STEL |
|---|---|---|---|---|---|---|
| Combustible (CH4) | 0–100 %LEL | 1 %LEL | 10 %LEL | 50 %LEL | — | — |
| Oxygen (O2) | 0–25 vol% | 0.1 vol% | 19.5 vol% low | 23.5 vol% high | — | — |
| Carbon monoxide | 0–150 ppm | 1 ppm | 25 ppm | 50 ppm | 25 ppm | 200 ppm |
| Hydrogen sulfide | 0–30 ppm | 0.5 ppm | 5.0 ppm | 30.0 ppm | 10.0 ppm | 15.0 ppm |
| Switch | Short press | Hold |
|---|---|---|
| POWER / ENTER | Confirm a menu item (starts a bump test) | 3 s — switch on or off |
| MODE | Step through PEAK / STEL / TWA / ALARM-P / BUMP / ID | — |
| RESET | Reset a latched alarm | Run the fresh air adjustment |
The status strip carries a pilot indicator, a carry state flag reading WORN / COVERED / NOT WORN, the battery icon and the clock. Where the aspirated instrument shows a pump indicator, this one shows whether it can actually breathe — the equivalent thing a wearer must keep an eye on.
A diffusion sensor needs air to reach it. Obstruct that and the instrument still displays a confident number; it is just wrong, and late.
| Carried | Reaching the sensor | Response | Verdict |
|---|---|---|---|
| Breathing zone (lapel/collar) | 100 % | Normal | Correct |
| Clipped at the belt | 92 % | 1.4× slower | Acceptable |
| Chest pocket, partly covered | 55 % | 3.2× slower | Restricted |
| In a toolbag | 18 % | 7× slower | Useless |
| Left outside the work area | 0 % | — | Monitoring nothing |
A bump test proves each channel responds to gas. It is not a calibration and adjusts nothing — which is precisely why it is the check you do daily while calibration is periodic. A channel passes if it reaches 50–150 % of the applied concentration; oxygen must actually see the depletion.
Keep it in the breathing zone and move through the work areas. Alarms latch: once a setpoint is reached the alarm stays up until you have genuinely left the gas. RESET pressed while still in the atmosphere is refused, and the simulator says so.
This is the part most often skipped and the reason the instrument exists.
Objective: prove the instrument before work. Expected result: fresh air adjust accepted, all four channels bump-pass, monitor worn in the breathing zone. Pass: 70. Par: 10 min.
Objective: work through the engine room (31 ppm CO) and hot work area (78 ppm CO), then review peak, STEL and TWA. Expected result: the STEL climbs sharply in the hot work area and the TWA keeps its value after you leave. Pass: 75. Par: 15 min.
Objective: the same work carried badly. Expected result: roughly a fifth of the true reading, arriving far too slowly; moving it to the breathing zone reveals the real atmosphere. Pass: 80. Par: 12 min.
Objective: an unventilated bilge at 17.9 vol% oxygen and 26 ppm H2S. Getting that reading requires standing in it. Expected result: recognise the alarm, leave, and answer the question in the control bar correctly — this instrument monitors your exposure; it does not clear a space for entry. Pass: 85. Par: 10 min.
Ctrl+Shift+F opens the fault panel: H2S sensor abnormality, unresponsive CO sensor (fails bump test), CO zero drift (+18 ppm), low battery, and a fouled sensor face that halves the response.
| Key | Action |
|---|---|
| F | Fullscreen |
| Space | Pause / resume |
| Esc | Close a modal |
| ? | Shortcut help |
| Ctrl+Shift+F | Fault injection panel |
| Ctrl+Shift+T | Trend plots |
| Ctrl+Shift+S | Score display |
| Event | Change |
|---|---|
| Bump test completed before use | +5 |
| Phase checklist completed | +5 |
| Correctly identifying the instrument's role | +15 |
| Alarm left unacknowledged over 30 s | −2 |
| Claiming a diffusion monitor can clear a space for entry | −25 |
Grades: A 90+, B 75–89, C 60–74, D below 60.
| Symptom | Cause | Action |
|---|---|---|
| Reading seems low for the area you are in | Sensor obstructed by carry position | Check the carry flag on the LCD; move it to the breathing zone |
| Reading changes very slowly | Covered sensor face, or a fouled face | Uncover it; inspect and clean the sensor face |
BUMP FAIL | A channel did not respond to test gas | Do not use the instrument. It cannot warn you on that channel |
FAIL CAL | Fresh air adjustment attempted in contaminated air | Move to genuinely clean air and repeat |
| RESET does nothing | The gas is still above the setpoint | Leave the area first — this is correct behaviour, not a fault |
Channel shows --- | Sensor failure | That gas is not being detected at all |
Reading shows ∩∩∩ | Above the service range | Leave immediately; the true value is unknown and higher |
| TWA will not go down | Working as designed | A dose already taken cannot be undone. Only a new reference period resets it |